| Model
SRH Specifications |
| Humidity Operating
Range |
0 to 100% RH |
| Accuracy (@23°C) |
2%, 3%, or 5% available |
| Long Term Stability |
<1%/Year @ 68°F
(20°C), 50% RH |
| User Selectable Electrical
Output: Voltage (3-Wire) |
0 to 5 VDC
0 to 10 VDC |
| Electrical Output:
Current (2-Wire) |
4 to 20 mA
|
| Excitation
|
13.5 to 30 VDC (10 VDC
Output)
12 to30 VDC (4-20 mA, 5 VDC Output) |
| Model Configuration |
Wall Mount
Duct Mount
Outside Air Mount |
| Operating
Temperature |
-40 to 140° F (-40 to
60° C) |
| Storage Temperature |
-40 to 158° F (-40 to
70° C) |
| Electrical Connection |
Pluggable Terminal Block |
|
Temperature Sensing Option |
|
Passive |
| T1: Thermistor |
NTC 10K
Ώ @ 77°F/25°C (Direct
Connect) Type II |
| T2: Passive RTD |
1000
Ώ @ 32°F/0°C (Direct Connect)
385 Platinum Curve |
|
Active |
| T3: Range
°F (°C) |
-58 to 140 (-50 to 60) |
|
Accuracy@ 68°F (20°C) Typ. @ 50% |
±1.1 (±0.6)
((Excitation 24 VDC +10%) |
| T5: Range
°F (°C) |
+14 to 140 (-10 to 60) |
|
Accuracy @ 68°F (20°C) Typ. @ 50% |
±0.7 (±0.4)
(Excitation 24 VDC +10%) |
| Signal Output Options |
Includes humidity
output |
| Current (2-Wire) |
4 to 20 mA |
| Field Selectable Voltage
(4-Wire) |
0 to 5 VDC, 0 to 10 VDC |
|
Replacement Sensor Assembly
P/N SRH3 |
| Accuracy |
2%, 3%, or 5% |
| Temperature Outputs |
None (RH Only)
10KΏ Thermister
(Passive)
1000Ώ RTD (Passive
-58 to 140°F (-50 to 60°C (Active))
+14 to 140°F (-10 to 60°C (Active)) |