The Mass Monitor Series consists of a moment-insensitive load cell module and circuit board. The load cell is available in any weighing capacity from 200g to 50kg, delivering weight measurements as acute as 2ppm with a linearity of 10ppm.
The circuit includes a display connector and a bidirectional RS-232 serial data interface. The self-contained load cell can easily be installed in a scale base or other weight-supporting structure, using only three bolts.
A display board using a six digit LED is available and may be used as an integral part of the Mass Monitor package or for testing purposes only.
Click here for the Mass Monitor
data sheet in PDF format
Requires
Adobe Acrobat Reader - Free Download here
| SETRA MODEL |
MASS MONITOR |
| Capacity x Readability:
|
| 200 |
g |
|
x |
|
.001 |
g |
| 1,000 |
g |
|
x |
|
.002 |
g |
| 2,500 |
g |
|
x |
|
.005 |
g |
| 5,000 |
g |
|
x |
|
.01 |
g |
| 10,000 |
g |
|
x |
|
.02 |
g |
| 25,000 |
g |
|
x |
|
.05 |
g |
| 50,000 |
g |
|
x |
|
.1 |
g |
|
| Weighing Range:
|
200 to 50,000 grams |
| Selectable Readability:
|
2, 5, 10+ ppm |
| Non-Linearity:
|
± 10 ppm |
| Hysteresis:
|
< ± 10 ppm |
| Creep:
|
< ± 10 ppm |
| Displacement:
|
0.01 in. to 0.05 in.*
(0.25 mm to 1.25 mm)
*Depends on range |
| Output:
|
RS-232 |
| Baud Rates:
|
300, 600, 1200, 2400, 4800, 9600 |
| Weight Update Rate:
|
0.2 sec. |
| Operating Temperature:
|
32°F to 122°F (0° to 50°C) |
| Operating Humidity:
|
0 to 90% R.H. |
| Display:
|
6 Digit LED |
| Power Requirement:
|
12 VDC @ 250 mA |
| Sensor Weight:
|
53 oz (1.5 Kg) |
| Sensor Dimensions:
|
7.5"l x 5"w x 3.5"h
19 cm x 12.5 cm x 9 cm
*Add 0.5" (1.5 cm) to height pan for load cell attachment posts |
| Warranty:
|
15 months from shipment |
| U.S. Patents:
|
The Mass Monitor series is covered by some or all of the following patents: 4,382,479; 4,383,586; 4,463,614; 4,448,085; 4,513,831; 4,558,600; 4,846,293 |
|
* Specifications subject to change without notice.
|
| |
|
©1998 Setra Systems, Inc. All rights reserved. All names and logos are trademarks. All other names, logos, and trademarks are property of their respective owners.
Web site design by DreamLight Incorporated: