Application
Designed to Weigh Gas Cylinders in Gas Cabinets and
Gas Distribution
|
| Model
CWS 321 Specifications |
| Platform Size |
9.25" (23.5cm) x 9.25" (23.5cm)
|
| Height |
0.97" (24.6mm)
|
| Platform Material |
Stainless Steel Pan |
| Weight |
10 lbs. (4.5kg) |
| Capacity |
0 to 60, 0 to 300, lbs
0 to 30, 0 to 135 kg |
| Electrical Termination |
10 ft. Cable or End
of Cable Bayonet Connector |
| Zero and Span
Adjustments |
External to Case |
| Performance
Data |
| Accuracy |
0.10% FS
|
| Non-Repeatability |
0.05% FS
|
| Hysteresis |
0.10%
FS
|
Off Center Loading
|
0.10% FS
(One half Scale load at 1 inch) |
| Overload Protection |
150% FS |
Thermal Effects
Compensated Range ºF (ºC)
Zero Shift %FS/100 ºF (50 ºC)
Span shift %FS/100 ºF (50 ºC) |
+14 to +122 (-10 to +50)
+/-1.0% (+/-0.9%)
+/-1.0% (+/-0.9%) |
| Long Term Stability |
0.5% FS/1YR |
| Environmental
Data |
Temperature
Operating ºF (ºC)
Storage ºF (ºC) |
+14 to +122 (-10 to +50)
+14 to +140 (-10 to +60) |
| Electrical
Data (Current Output) |
| Circuit |
2-wire |
| Output |
4-20 mA |
| External Load |
0 to 800
Ohms |
Minimum supply
voltage (VDC) = 15 + 0.02 x (Resistance of
receiver plus line).
Minimum supply voltage (VDC) = 30 + 0.02 x (Resistance
of receiver plus line).
A regulated power
supply or a power supply with low ripple noise is
recommended for best performance. |
Click here for data sheet Model CWS 321
* Specifications subject to change
without notice. If you don't see what you need here,
please contact us.
|