Model 516 OEM
Gauge and Compound
Pressure
|
|
Applications
- General Purpose
- Industrial OEM Equipment
- Off-Highway Vehicles
- Pumps and Compressors
- Process Systems
- Hydraulic Systems
- Industrial Engines
|
| Model
512 Specifications
|
| Electrical Output Voltage
|
0 to 5 VDC, 0 to 10 VDC, 0.5 to 5.5
VDC, 1 to 5 VDC, 1 to 6 VDC, 1 to 11 VDC |
| Excitation
|
1.5 VDC Above Span to 35 VDC @ 6 mA |
| Electrical Output Current
|
4 to 20 mA |
| Loop Supply Voltage
|
24 VDC (7-35 VDC) |
Accuracy
(RSS Method)
|
±0.5% Full Scale |
| Type of Pressure
|
Gauge
Compound |
| Pressure Ranges
|
0 to 15, 30, 60, 100,150, 200, 300, 500, 600, 1000,
1500, 2000, 3000, 4000, 5000, 6000 psig
14.7 to 0, 15, 45, 135, 185, 285 psig
1, 1.6, 2.5 4.0, 6.0, 10, 16, 25, 40, 60, 100, 160, 250, 400 bar |
| Thermal Effects
|
- Compensated Range °F (°C):
- -5 to +180 (-20 to +80)
- %FS/100°F(100°C)max.zero:
- ±1.0 (±2.0)
- %FS/100°F(100°C)max.span:
- ±1.0 (±2.0)
|
| Media
|
Gases or liquids compatible with 17-4PH stainless steel. (Hydrogen not recommended for use with 17-4PH stainless steel) |
| Option
|
Intrinsic Safe (ETL approved for Class 1, Div 1.
Groups C & D, hazardous areas) |
Click here for data sheet:
Model 516
* Specifications subject to change without notice. If you don't see what you need here, please contact us.
|
| |
|
©1998 Setra Systems, Inc. All rights reserved. All names and logos are trademarks. All other names, logos, and trademarks are property of their respective owners.