Model 204/C204

High accuracy for corrosive liquids or gases - gauge, vacuum and absolute pressures

Click here for product data sheet:   Model 204 Data Sheet

Email to a Friend

Customers also bought:

  • High accuracy pressure for corrosive liquids or gases – gauge, compound and absolute

  • Meter and manometer, analog to digital converter w/ LED display, optional transducer

 Overview
Ideally suited to applications that require 0.1% accuracy. The Model 204/C204 is versatile in its use and is well suited for critical pressure measurement (gage, absolute, vacuum) in Laboratory, Industrial and Aerospace applications.

Available in optional FM Approved Intrinsically Safe design.
Features
> Fast Warm-Up

> 0-5 VDC or 4-20 mA Output

> 0.11% FS Accuracy

> Excellent Thermal Effects

> Low Output Noise

> Fast Response, Less than 1 Millisecond

> Solid One-Piece Stainless Steel Sensor

>FM Approved Intrinsically Safe (Option)

> Meets CE Conformance

Pressure Ranges

0 to 25, 50, 100, 250, 500, 1000, 3000, 5000, 10000 psig
0 to 25, 50, 100, 250, 500, 1000, 3000, 5000 psia
0 to 14.7 psiv

Accuracy

±0.11% FS
±0.14% for 10,000 PSIG
 
Output

0 to 5 VDC
24 VAC
 
Excitation

Voltage:  22 to 30 VDC
Current:  24V (DC Only)

Media
Gases or liquids compatible with 17-4 PH stainless steel.

Electrical Options Model 204 Only

602-606: Special Outputs: 1 to 5, 1 to 6, 0 to 10, 1 to 10,
0 to 2.5 VDC.

621: Special Excitation: ±24 VDC

622: Special Excitation: ±15 VDC
640: Intrinsic Safe Design FM Approved for class
I, II,III, Division 1, Groups A, B, C, D, E, F, G
hazardous locations. For standard catalog
products without other electrical options or remote controls options

642: Remote Sensitivity (Not Available with Options 621, 622)

643: Remote Calibration (adjustable)

644: Remote Calibration (fixed)
 
Performance Options Model 204 or C204

702: Compensated Temperature, -65°F to +250°F;
(Doubled the Standard Thermal Effect
Specification)

710: 0.073% FS (RSS) Accuracy (not available on
10,000 PSIG ranges)
 
Mechanical Options Model 204 or C204

803-825: Up to 25 ft. of cable can be supplied on your
order; please specify cable length when ordering
(eg. 805 for 5ft. cable). Consult factory for cables
longer than 25 ft

865 NEMA 4 Weather Proof Enclosure

866 Explosion Proof PT Housing, FM approved for:
CL. I, Div 1 & 2, GP. B, C, D; CL. II. Div 1 &2,
GP. E, F, G; CL. III, Div 1&2 and NEMA 4 weather
proof when factory installed in this enclosure

904: Cleaning for Oxygen Service

911: Etched Metal Stainless Steel Tags
 Documents and Downloads